Friday, 13 September 2024
Product Detail of - DIGITAL MULTIMETER DT4256
Product Category : Digital Multimeters
Series : High Precision Handheld DMM

Fully Loaded DMM with 11 Functions For General Purpose Electrical Testing.

The Hioki DT4200 Digital Multimeter Series consist of a complete lineup of 9 models ranging from professional to industrial to pocket meters, delivering a super fast response rate and safety features that take electrical testing to a higher level. The DT4256 is a digital multimeter designed to deliver maximum functionality for use in a wide range of settings, including instrumentation testing with a 4-20 mA% conversion function.

Feature:

• 11 functions including 4-20 mA% conversion
• ±0.3% DC V basic accuracy, wide 40 Hz to 1 kHz AC V frequency characteristics
• Low-pass filter (100 Hz/ 500 Hz) cuts high harmonics(when measuring inverter fundamental waveforms)
• Connect to PC with USB communication function (optional)
• Broad -10 (14°F) to 50°C (122°F) operating temperature range

Specification:

■ Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Regarding DMM Accuracy
Due to the many ranges and functions available in a DMM, only the basic accuracy is indicated for reference. Please refer to the individual catalogs for detailed accuracy information.

DC Voltage (Basic accuracy) 600.0 mV to 1000 V
5 ranges, Basic accuracy: ±0.3 % rdg. ±3 dgt.
AC Voltage 40 to 500 Hz 6.000 V to 1000 V, 4 ranges, Frequency characteristics: 40 Hz to 1 kHz
Basic accuracy: ±0.9 % rdg. ±3 dgt. (True RMS, crest factor 3)
AUTO AC/DCV Yes
Resistance 600.0 Ω to 60.00 MΩ, 6 ranges, Basic accuracy: ±0.7 % rdg. ±3 dgt.
DC Current Basic accuracy 60.00 mA to 10.00 A, 4 ranges, ±0.9 % rdg. ±3 dgt.
From 4 to 20 mA% conversion display Yes
AC Current (direct input) 40 to 500 Hz Basic accuracy 600.0 mA to 10.00 A, 3 ranges,
±1.4 % rdg. ±3 dgt.
(True RMS, crest factor 3, 40 Hz to 1kHz)
AC Current (use with Clamp on probes) 40 to 1k Hz 10.00 A to 1000 A, 7 ranges,
Add the clamp on probe accuracy to basic accuracy: ±0.9 % rdg. ±3 dgt. 
(True RMS, crest factor 3)
Voltage detection Hi: AC40 V to 600 V, Lo: AC80 V to 600 V
C (capacitance) 1.000 μF to 10.00 mF, 5 ranges, Basic accuracy: ±1.9 % rdg. ±5 dgt.
Frequency 99.99 Hz (5 Hz or more) to 99.99 kHz, 4 ranges (limited by the minimum detectable voltage and current), 
Basic accuracy: ±0.1 % rdg. ±1 dgt.
Continuity check Continuity threshold [ON]: 25 Ω or less (Indicate buzzer sound, red LED), 
[OFF]: 245 Ω or more, Response time: 0.5 ms or more
Diode test Open terminal voltage: 5.0 V or less, Testing current 0.5 mA or less, 
Threshold of forward voltage: 0.15 V to 1.5 V
Other functions Filter function, Display value hold, Auto hold, Max/Min/Average value display, Relative display, Auto-power save, USB communication (option)
Display Main and Sub displays: 4-digits LCD, max. 6000 digits, bar graph
Display refresh rates 5 times/s (Capacitance measurement: 0.05 to 5 times/s, depending on measured value, Frequency: 1 to 2 time/s, Temperature: 1 time/s )
Power supply LR03 alkaline batteries ×4, Continuous use: 130 hours (backlight OFF)
Dimensions and mass 84 mm (3.31 in)W × 174 mm (6.85 in)H× 52 mm (2.05 in)D,
390 g (13.8 oz) (including batteries and holster)
Accessories Test lead L9207-10 ×1, Holster ×1, Instruction manual ×1, LR03 alkaline battery ×4

 

Back